Table 4.
Electrophoresis Conditions
| Condition | Capillary | Polymer/urea | Voltage (V/cm) | Temperature (°C) | Run time (min) |
| GSP1 | Lt 41 cm | 3%/6.6 m | 317 | 30 | 21 |
| Ld 30 cm | |||||
| i.d. 75 μm | |||||
| GSP2 | Lt 61 cm | 3%/8 m | 246 | 30 | 41 |
| Ld 50 cm | |||||
| i.d. 75 μm | |||||
| POP-4a | Lt 41 cm | POP-4 | 366 | 60 | 23 |
| Ld 30 cm | |||||
| i.d. 50 μm | |||||
| POP-4b | Lt 47 cm | POP-4 | 319 | 60 | 31 |
| Ld 36 cm | |||||
| i.d. 50 μm |
-
↵(Lt) Total length of capillary; (Ld) length of capillary from injection to detector window; (i.d.) internal diameter.
-
↵Run time includes electrophoresis time plus the times required to refill the capillary after each analysis with fresh polymer, sample injection, and rinsing of the capillary/electrode tip.