ACAPELLA-1K, A Capillary-Based Submicroliter Automated Fluid Handling System for Genome Analysis

Table 1.

Summary of ACAPELLA-1K Processing Results

Subsystem/assay No. of samples Failure rate
Total capillaries 27,000 all tests for year
Capillary dispenser (hopper) >8,000 0.063%
DNA sample aspirator >8,000 0%
Transport and offloader >8,000 4% improved to 0.1%
Reagent dispensers >8,000 0.1% improved to 0.056%
Mixer >8,000 0% (>8000 capillaries)
Capillary cassettes during thermal cycling >8,000 2.9% improved to 0% with addition of clip
PCR results >9,000 0%–4%, 1.57% average
Sequencing reactions >200 0%
  • Results are from September 1998 to August 1999.