RIKEN Integrated Sequence Analysis (RISA) System—384-Format Sequencing Pipeline with 384 Multicapillary Sequencer

Table 1.

Comparison of a Scanning System and An Image Detection System

Detectable photon number/unit time (Dpnt) Detection time (Dt) Total detectable photon number from a single capillary during a single sampling interval (Dpn)
Scanning system A × Np C × SI/N A × C × Np × SI/N
Image detection system Np/N2 SI × RT × N Np × (SI/N2 − RT/N)
Ratio A × C × (1/N − RT/SI) ≈ A × C/N
  • (A, C) Constants depending on the scanning optical and mechanical system; (Np) detectable photon number/unit time for static single capillary detection system; (SI) sampling interval; (N) number of capillaries; (RT) reading time of total data of the CCD element.